1 /2





























9000 SERIES     9100 SERIES     9500 SERIES     9600 SERIES     9700 SERIES     

Programmable Model 9512
The Programmable Controller offers a wide temperature range, exceptional stability, and variable-speed pressure/suction (duplex) pump along with time/temperature programming and remote probe capability. RS232 interface and PC programming software are standard. LabView™ drivers and Excel® macros provide even greater programming and data logging convenience. The full graphic LCD display and multi-language help menus simplify operation and set-up.

Advanced Digital Model 9502
Excellent temperature range and stability combined with an easy-reading graphic display make this model a fine choice for many applications. The pump is a powerful variable-speed pressure/suction (duplex) type for open or closed loop applications. A built-in RS232 interface comes standard. The full graphic LCD display and multi-language help menus simplify operation and set-up.

Standard Digital Model 9506
The same -30ºC cooling capacity of our Programmable and Digital models at a price comparable to units with analog controllers make the Model 9506 a great value for applications where high heat removal at close to ambient temperature is required. Microprocessor-based controller provides set-and-read functionality; three user-defined temperature buttons make set point changes quick and easy. The
2-speed pressure (simplex) pump is suitable for closed loop applications.

 

Cooling
Heating

Time in Minutes

 

9500 Series Specifications
Reservoir Capacity
13 liters
Cooling Capacity @

+20ºC660 Watts
0ºC480 Watts
-20ºC240 Watts
-30ºC100 Watts

Over-Temperature Cutoff
Adjustable
Reservoir Drain
Yes
Low-Liquid Cutoff
Yes
Working Access
l x w x h
6 x 11 x 5 1/2 in.
14 x 28 x 15.2 cm
Overall Dimensions
l x w x h
17 x 15 1/2 x 24 3/4 in.
43 x 39.4 x 62.9 cm
Pump Inlet & Outlet
1/4 in. FPT Rear Discharge

 

 

 

FEATURES

-30ºC Cooling Capacity

13 Liter Reservoir

Energy-Saving Cool Command™ Refrigeration System

Large Reservoir, Large Opening

Easy-Access Reusable Filter

CFC-Free Refrigerant



PolyScience 9500 Series Circulators offer exceptional cooling capability over a broad temperature range plus a high-capacity, easy-to-access 13 liter reservoir.

These versatile circulators also feature our exclusive Cool Command refrigeration system and fuzzy logic energy management technique to deliver up to 50% greater energy efficiency than traditional systems. This advanced technology carefully monitors heat load to deliver only the amount of refrigerant needed. As a result, it permits fast bath cooling even at high temperatures without excessive energy consumption.

To further add to their application versatility, Models 9512 and 9502 feature a powerful 5-speed pressure/suction (duplex) pump, making them ideal for use where continuous, low-temperature operation is required or where the viscosity of the liquid may affect circulation uniformity.



APPLICATIONS

Low-temperature
Calibration

Cloud & pour point
testing

Cell freezing

Low-temperature
Reactions

Distillation Condensers

General Laboratory
Cooling

Rotary Evaporators

Spectrophotometers

Viscosity Studies

 
Specification Model 9512 Model 9502 Model 9506
Temperature Range -30º to 200ºC -30º to 200ºC -30º to 150ºC
Temperature Stability +/- 0.01ºC +/- 0.01ºC +/- 0.05ºC
Controller/RS232 Programmable/Yes Advanced Digital/Yes Standard Digital/No
Readout Graphic LCD Graphic LCD LED
Temperature Readout ºC or ºF ºC or ºF ºC or ºF
Readout Accuracy +/- 0.25ºC +/-0.25ºC +/- 0.5ºC
Pressure Flow Rate Variable-Speed
30 lpm max. (60Hz)
22 lpm max. (50Hz)
Variable-Speed
30 lpm max. (60Hz)
22 lpm max. (50Hz)
2-speed
5 or 12 liters/min
Suction Flow Rate Variable-Speed
22 lpm max. (60Hz)
15 lpm max. (50Hz)
Variable-Speed
22 lpm max. (60Hz)
15 lpm max. (50Hz)
N/A
Heater (60Hz/50Hz) 1100/2200 Watts 1100/2200 Watts 1100/1600 Watts
Shipping Weight 139 lbs. / 63.2 kg 139 lbs. / 63.2 kg 139 lbs. / 63.2 kg
Cat. No. 120V/60Hz/13A 9512A11C 9502A11C 9506A11C
Cat. No. 240V/50Hz/9.8A 9512A12E 9502A12E 9506A12E


Notes: Cooling capacity, Watts x 3.41 = BTU's/hr. Performance specifications determined at ambient temperature of 20ºC/68ºF. For 50Hz units, derate cooling capacity and pump flow by 17%.