ACCESSORIES

 

 

PolyScience KR and VLT Series Coolers can be used to achieve low-
temperature capability with non-refrigerated circulators or boost the cooling capacity of refrigerated circulators. They're also ideal for replacing dry ice or liquid nitrogen. All models in the KR and VLT Series are designed to run at maximum cooling rate; temperature control must be provided by an external circulator.

Immersion Probe style coolers
are excellent for trapping and dewar type applications and can be used to reduce the expense of using dry ice or liquid nitrogen. A 3-ft. (.9 m) flexible hose allows convenient placement of the cooling probe.

Flow-Through style coolers
must be used in-line with a circulator pump. Because of their high cooling efficiency, an anti-freeze solution is required in the circulating system.

 

FEATURES

VLT Series Models Now Available With Digital Set/Read Controller

Compact Package

Low-Cost Operation

Designed For
Continuous Use

Replaces Dry Ice
or Liquid Nitrogen



APPLICATIONS

Cooling Exothermic
Reactions

Freezing Point
Determinations

Impact Testing

Lyophilization

Quick Cooling

Replace Dry Ice

Solvent Traps

Vacuum Traps

 

 

 

 

Immersion Probe KR 50 KR 80
Operating Range -35º to 40ºC -45º to 40ºC

Cooling Capacity at -30ºC
Cooling Capacity at 20ºC

140 Watts
975 Watts
265 Watts
1050 Watts
Probe Coil
Diameter x Length
1 3/4 dia. x 4 in.
3.8 dia x 10.2 cm
3 dia. x 9¼ in.
7.6 dia. x 23.5 cm

Overall Dimensions
l x w x h

17 x 14 x 14 in.
43.2 x 35.6 x 35.6 cm
17 x 14 x 14 in.
43.2 x 35.6 x 35.6 cm
Shipping Weight 76 lbs. / 34.5 kg 82 lbs. / 37.7 kg
Cat. No. 120V/60Hz, 5A 070850 070880
Cat. No. 240V/50Hz, 2.5A 070855 070885

 

Flow-Through KR 60
Operating Range -25º to 40 ºC

Cooling Capacity at -10ºC
Cooling Capacity at 20ºC

260 Watts
745 Watts
Inlet / Outlet size 3/8 in. FPT

Overall Dimensions
l x w x h

17 x 14 x 14 in.
43.2 x 35.6 x 35.6 cm
Shipping Weight 78 lbs. / 35.5 kg
Cat. No. 120V/60Hz, 5A 070860
Cat. No. 240V/50Hz, 2.5A 070865

 

Specifications VLT 100-NC VLT 60
Operating Range -40ºC to -100ºC -20ºC to -60ºC
Temperature Display LED N/A
Compressor Two; 1/4 HP each one; 1/4 HP

Cooling Capacity:

150 Watts @ -40ºC
120 Watts @ -60ºC
70 Watts @ -80ºC
75 Watts @ -20ºC
0 Watts @ -60ºC
Standard Cooling Probe
3 in. Ridgid Coil
1.5 in. Rigid Coil
Optional Cooling Probes Cold Finger
Flexible
None
Overall Dimensions
h x w x d
21 x 15 x 20.125 in.
53.3 x 38.1 x 51.1 cm
9 x 10 x 11 in.
22.9 x 25.4 x 27.9 cm
Shipping Weight 185 lbs. / 84.2 kg 49 lbs. / 22.3 kg
Electrical 120V, 50/60 Hz, 12A
240V, 50/60 Hz, 6A
120V, 60 Hz, 5A
240V, 50 Hz, 2.5A

Probe Style / Cat. No.
(120V)

3 In. Rigid Coil / P10N6A101B
Cold Finger / P10N3A101B
Flexible / P10N4A101B
1.5 In. Rigid Coil / 070960
Probe Style / Cat. No
(240V)
3 In. Rigid Coil / P10N6A102E
Cold Finger / P10N6A102E
Flexible / P10N4A102E
1.5 In. Rigid Coil / 070965

 

  Insulated Flexible Hose 3 In. Rigid Coil Probe
(VLT 100, KR 80)
1.75 In. Rigid Coil Probe
(VLT 60)
Diameter VLT 100 - 2.83 in. (7.14 cm)
VLT 60 - 1.5 in. (3.81 cm)
KR Series - 1.5 in. (3.81cm)
3 in. (7.62 cm)

1.75 in. (4.44 cm)

Length VLT 100 - 6 ft. (1.83 m)
VLT 60 - 4 ft. (1.22 cm)
KR Series - 4 ft. (1.22 m)
15 in. (38.1 cm)
exposed length;
8.75 in. (22.2 cm)
Coil length
15 in. (38.1 cm)
exposed length;
7.5 in. (19.05 cm)
Coil length

 

  1.5 In. Rigid Coil Probe
(VLT 60)
Cold FInger Probe
(VLT 100)

Flexible Probe
(VLT 100)

Diameter

1.5 in. (3.81 cm)

0.75 in. (1.91 cm) 0.625 in. (1.59)
Length 15 in. (38.1 cm)
exposed length;
4.5 in. (11.4 cm)
Coil length
12 in. (30.48 cm) 15 in. (38.1 cm)