Model 4100 Liquid-to-Liquid Cooler

Liquid-to-Liquid Cooler

Select voltage
$4,199.00
4150T2

Key Specifications

Description

Liquid-to-Liquid Cooler

Connectivity
Optional: RS232, RS485, Remote On/Off
Working Temperature Range °C
Facility water +10° to 60°
Temperature Stability °C
±0.4°
Temperature Calibration Capability
1-point
Pump Type
Turbine
External Temperature Control Capability
Optional

Operator's Manual

Key Features

  • Protects precision equipment from facility water contaminants
  • Quiet, energy-efficient liquid-to-liquid heat removal
  • Displays temperature and pressure or flow rate information
  • Built-in temperature and low flow alarms

Common Specifications

Working Temperature Range °F
Facility water +20° to 140°
Working Temperature Range °C
Facility water +10° to 60°
Reservoir Capacity (gallons)
1.11
Reservoir Capacity (liters)
4.2
Reservoir Cover
Screw-on Cap
Cleanable Fluid Filter
Yes
Temperature Stability °F
±0.8°
Temperature Stability °C
±0.4°
Pump Type
Turbine
Pump Speed
Constant, Single Speed
Process Connections
1/2" (F) NPT
External Temperature Control Capability
Optional
Over-Temperature Protection / Failsafe Heater Control
Yes
Reservoir Drain
Yes
Maximum Ambient Temperature °F
95°
Maximum Ambient Temperature °C
35°
Overall Dimensions (L x W x H) (inches)
27.6 x 14.5 x 22.6 in
Overall Dimensions (L x W x H) (cm)
70.2 x 36.8 x 57.5 cm
Shipping Weight (pounds)
175.0
Shipping Weight (kilograms)
79.5
Catalog Page Number
108

60 Hz Only

Part Number
4150T21A130D
Maximum Pressure (psi)
100.0
Maximum Pressure (bar)
6.90
Maximum Pressure Flow Rate (gpm)
3.50
Maximum Pressure Flow Rate (l/min)
13.2
Cooling Capacity (W/BTU)
15000 / 52855 based on 15°C ΔT
10000 / 34100 based on 10°C ΔT
4500 / 5345 based on 5°C ΔT
ΔT = Process water temperature - facility water temperature
Included Hardware
Two sets of Inlet/Outlet Adapters: ½ inch male NPT, 5/8 inch male NPT
Electrical Requirements (VAC/Hz/Ph/A)
208-230/60/1/3
Regulatory Approvals
TUV

50 Hz Only

Part Number
4150T21A130E
Maximum Pressure (psi)
83.0
Maximum Pressure (bar)
5.70
Maximum Pressure Flow Rate (gpm)
2.90
Maximum Pressure Flow Rate (l/min)
11.0
Cooling Capacity (W/BTU)
15000 / 52855 based on 15°C ΔT
10000 / 34100 based on 10°C ΔT
4500 / 5345 based on 5°C ΔT
ΔT = Process water temperature - facility water temperature
Included Hardware
Two sets of Inlet/Outlet Adapters: ½ inch male NPT, 5/8 inch male NPT
Electrical Requirements (VAC/Hz/Ph/A)
240/50/1/3
Regulatory Approvals
CE