Operator's Manual
Key Features
- Continuous cooling to temperatures as low as -100°C
- Designed to run at maximum cooling
- An economical alternative to dry ice or liquid nitrogen
- Excellent for trapping applications, freeze drying, and rapidly cooling small volumes of liquids
Common Specifications
Working Temperature Range °F |
-112° to -40° |
---|---|
Working Temperature Range °C |
-80° to -40° |
Temperature Control (°F) |
Fixed at -112° |
Temperature Control (°C) |
Fixed at -80°C |
Cleanable Air Filter |
Yes |
Display |
Temperature |
Maximum Ambient Temperature °F |
86° |
Maximum Ambient Temperature °C |
30° |
Overall Dimensions (L x W x H) (inches) |
20.1 x 15 x 22.3 in |
Overall Dimensions (L x W x H) (cm) |
51.1 x 38.1 x 56.6 cm |
Refrigerant |
R404A & R508B |
Immersion Probe Type |
Rigid Coil |
Probe Diameter (inches) |
1.875 |
Probe Diameter (cm) |
4.76 |
Probe Length (inches) |
Coil: 7 / Exposed: 16.5 |
Probe Length (cm) |
Coil: 17.8 / Exposed: 41.9 |
Flexible Hose Diameter (inches) |
2.83 |
Flexible Hose Diameter (cm) |
7.14 |
Flexible Hose Length (feet) |
6 |
Flexible Hose Length (m) |
1.83 |
Shipping Weight (pounds) |
162.0 |
Shipping Weight (kilograms) |
73.5 |
Catalog Page Number |
114-115 |
60 Hz Only
Part Number |
P80NHA101B |
---|---|
Cooling Capacity @ -80°C (W) |
100 |
Cooling Capacity @ -60°C (W) |
215 |
Electrical Requirements (VAC/Hz/Ph/A) |
120/60/1/11.5 |
50 Hz Only
Part Number |
P80NHA102E |
---|---|
Cooling Capacity @ -80°C (W) |
100 |
Cooling Capacity @ -60°C (W) |
215 |
Electrical Requirements (VAC/Hz/Ph/A) |
240/50/1/7.5 |
Regulatory Approvals |
CE |