00:00

Heated Circulating Baths

Integrated Heated Circulating Baths with
Advanced Digital Temperature Controller

Key Specifications

Working Temperature: Ambient +10° to +200°C 
Temperature Stability: ±0.01°C 
Safety Rating: DIN 12876-1 Class III (for use with flammable liquids) 
External Temperature Control Capability: Yes 
Pump: Variable-speed 
External Circulation: Open- or closed-loop 

  • Integral tap water cooling coil
  • LidDock™ self-stowing reservoir cover
  • DuraTop™ insulating, chemical-resistant top deck
  • Reservoir drain

Key Features

  • Intuitive 3.75” (9.5 cm) display with touch-pad control
  • 4 languages: French, German, Spanish, English
  • Swivel 180™ Rotating Controller
  • On-board connectivity: USB-A & B, Ethernet, RS-232/RS-485 and external temperature probe
  • On-screen prompts
  • Automatic performance optimization and specific heat tuning
  • Single-point calibration capability

Performance Specifications & Part Numbers

7 Liter Heated Circulator 15 Liter Heated Circulator 20 Liter Heated Circulator 28 Liter Heated Circulator
Working Temperature Range Ambient +10° to 200° C Ambient +10° to 200° C Ambient +10° to 200° C Ambient +10° to 200° C
Heater Wattage 1100 W (60Hz)
2200 W (50Hz)
1100 W (60Hz)
2200 W (50Hz)
1100 W (60Hz)
2200 W (50Hz)
1100 W (60Hz)
2200 W (50Hz)
60Hz Part Number
AD07H200-A11B
AD15H200-A11B
AD20H200-A11B
AD28H200-A11B
50Hz Part Number AD07H200-A12E AD15H200-A12E AD20H200-A12E AD28H200-A12E

Optimized

The Advanced Digital Temperature Controller displays and controls bath temperatures with ±0.01°C precision and stability. Plus you can further optimize performance through a variety of control settings including pump speed, specific heat value, calibration offset and internal or external temperature control. 

Communicative

Bi-directional RS-232, addressable RS-485 and Ethernet communications provide fast and versatile remote control and data logging. USB-A and USB-B enable you to log data directly to a flash drive or computer.

Friendly

On-screen prompts and 4 languages (English, Spanish, French and German) take the guesswork out of setup and operation. Adjusting the temperature set-point, pump speed or other operational settings is simple, straightforward and intuitive. In addition, the Advanced Digital Temperature Controller features a message bar that assists in making changes and provides alerts and safety information.

Safe

The Advanced Digital Temperature Controller delivers exceptional safety for a wide variety of laboratory applications. Standard safety features include redundant over-temperature protection, low-liquid level float switch, user-settable over-temperature safety and high and low temperature limits.